Human Metallothionein 1 (MT-1) ELISA Kit

Alternative:MT1A; MT1S; MTC
Detection Range:0.31-20ng/mL
Sensitivity:0.11ng/mL
Assay Type:Sandwich

Product Name:

Human Metallothionein 1 (MT-1) ELISA Kit

Cat.#:

KB32672

Regulatory Status:

RUO

Alternative:

MT1A; MT1S; MTC

Detection Method:

ELISA

Assay Type:

Sandwich

Detection Range:

0.31-20ng/mL

Sensitivity:

0.11ng/mL

Specificity:

This kit recognizes Human MT-1 in samples. No significant cross-reactivity or interference between Human MT-1 and analogues was observed.

Sample Type:

Serum, plasma, tissue homogenates and other biological fluids

Research Area:

Cancer


| MT-1

Metallothionein (MT) is a family of cysteine-rich, low molecular weight (MW ranging from 500 to 14000 Da) Proteins. They are localized to the membrane of the Golgi apparatus. MTs have the capacity to bind both physiological (such as zinc, copper, selenium) and xenobiotic (such as cadmium, mercury, silver, arsenic, lead) heavy metals through the thiol group of its cysteine residues, which represent nearly 30% of its constituent amino acid residues.There are four main isoforms expressed (family 1, see chart below): MT1 (subtypes A, B, E, F, G, H, L, M, X), MT2, MT3, and MT4.


| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

  

Concentration (ng/mL)201052.51.250.620.310
OD2.61.91.240.750.460.310.220.07
Corrected OD2.531.831.170.680.390.240.15-

 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human MT-1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 

Sample TypeRange(%)Average Recovery(%)
Serum(n=8)84-10196
Plasma(n=8)92-105102
Tissue Homogenates(n=8)96-108105






 

| Linearity

Samples were spiked with high concentrations of Human MT-1 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 



Serum(n=4)Plasma(n=4)Tissue Homogenates(n=4)
1 : 2Range (%)84-9588-9790-110
Average (%)919396
1 : 4Range (%)89-10387-108105-115
Average (%)9498108







Inquiry

Scan the qr codeclose
the qr code